Abstract

Purpose

Exploring the relationship between the signal-to-noise ratio (SNR) of organs and size-specific dose estimate (SSDE) in tube current modulation (TCM) chest CT examination.

Methods

Forty patients who received TCM chest CT scanning were retrospectively collected and divided into four groups according to the tube voltage and sexes. We chose to set up the region of interest (ROI) at the tracheal bifurcation and its upper and lower parts in slice images of the heart, aorta, lungs, paracranial muscles, and female breast, and the SNR of each organ was calculated. We also calculated the corresponding axial volume CT dose index (CTDIvolz) and axial size-specific dose estimate (SSDEz).

Results

The correlation analysis showed that the correlation between the SNR of the slice images of most organs and SSDEz was more significant than 0.8, and that between the SNR and CTDIvol was more significant than 0.7. The simple linear regression analysis results showed that when the sex is the same, the SNR of the same organ at 100kVp was higher than 120kVp, except for the lung. In multiple regression analysis, the result indicated that the determination coefficients of the SNR and SSDEz of the four groups were 0.934, 0.971, 0.905, and 0.709, respectively.

Conclusion

In chest CT examinations with TCM, the correlation between the SNR of each organ in slice images and SSDEz was better than that of CTDIvolz. And when the SSDEz was the same, the SNR at 100 kVp was better than that at 120 kVp.

This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International Public License (CC-BY 4.0), a copy of which is available at: https://creativecommons.org/licenses/by/4.0/legalcode. This license permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Loading

Article metrics loading...

/content/journals/cmir/10.2174/0115734056267251231121064259
2024-02-20
2024-12-27
Loading full text...

Full text loading...

/deliver/fulltext/cmir/20/1/CMIR-20-E15734056267251.html?itemId=/content/journals/cmir/10.2174/0115734056267251231121064259&mimeType=html&fmt=ahah

References

  1. DamilakisJ. DosimetryC.T. CT Dosimetry.Invest. Radiol.2021561626810.1097/RLI.000000000000072732932380
    [Google Scholar]
  2. ShopeT.B. GagneR.M. JohnsonG.C. A method for describing the doses delivered by transmission x-ray computed tomography.Med. Phys.19818448849510.1118/1.5949957322067
    [Google Scholar]
  3. SteinigerB. KlippelC. TeichgräberU. ReichenbachJ.R. FiebichM. Can the size-specific dose estimate be derived from the body mass index? a feasibility study.Radiat. Prot. Dosimetry2022198632533310.1093/rpd/ncac03835443046
    [Google Scholar]
  4. StraussK. BooneJ. CodyD. McColloughC. McNitt-GrayM. TothT. Size Specific Dose Estimates in Pediatric and Adult Body CT Examinations,AAPM Task Group20142011
    [Google Scholar]
  5. BooneJ.M. StraussK. CodyD.D. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations.(AAPM)College Park (MD): American Association of Physicists in Medicine2011
    [Google Scholar]
  6. YuanZ. Comparison of different SSDE methods in evaluating the radiation dose in chest and abdomen CT scan.Chi. J. Radi. and Prot.20163611852856
    [Google Scholar]
  7. McColloughC.H. BruesewitzM.R. KoflerJ.M.Jr CT dose reduction and dose management tools: Overview of available options.Radiographics200626250351210.1148/rg.26205513816549613
    [Google Scholar]
  8. KwonJ.K. ChangI.H. MoonY.T. LeeJ.B. ParkH.J. ParkS.B. Usefulness of low-dose nonenhanced computed tomography with iterative reconstruction for evaluation of urolithiasis: diagnostic performance and agreement between the urologist and the radiologist.Urology201585353153810.1016/j.urology.2014.11.02125733262
    [Google Scholar]
  9. LarsonD.B. Optimizing CT radiation dose based on patient size and image quality: The size-specific dose estimate method.Pediatr. Radiol.201444S3Suppl. 350150510.1007/s00247‑014‑3077‑y25304711
    [Google Scholar]
  10. KhoramianD. SistaniS. FirouzjahR.A. Assessment and comparison of radiation dose and image quality in multi-detector CT scanners in non-contrast head and neck examinations.Pol. J. Radiol.201984616710.5114/pjr.2019.8274331019596
    [Google Scholar]
  11. SpampinatoM.V. StalcupS. MatheusM.G. ByingtonK. TylerM. BickleyS. TipnisS. Radiation dose and image quality in pediatric head ct.Radiat. Prot. Dosimetry2018182331031610.1093/rpd/ncy06630590841
    [Google Scholar]
  12. RajaramanV. PonnusamyM. HalanaikD. Size specific dose estimate (SSDE) for estimating patient dose from CT used in myocardial perfusion SPECT/CT.Asia Ocean. J. Nucl. Med. Biol.202081586332064284
    [Google Scholar]
  13. BarretoI. VermaN. QuailsN. OlguinC. CorreaN. MohammedT.L. Patient size matters: Effect of tube current modulation on size-specific dose estimates (SSDE) and image quality in low-dose lung cancer screening CT.J. Appl. Clin. Med. Phys.2020214879410.1002/acm2.1285732250062
    [Google Scholar]
  14. YurtA. Özsoykalİ. ObuzF. Effects of the use of automatic tube current modulation on patient dose and image quality in computed tomography.Mol. Imaging Radionucl. Ther.20192839610310.4274/mirt.galenos.2019.8372331507141
    [Google Scholar]
  15. LarsonD.B. MalarikR.J. HallS.M. PodbereskyD.J. System for verifiable CT radiation dose optimization based on image quality. part II. process control system.Radiology2013269117718510.1148/radiol.1312232123784877
    [Google Scholar]
  16. BraunF.M. JohnsonT.R.C. SommerW.H. ThierfelderK.M. MeinelF.G. Chest CT using spectral filtration: Radiation dose, image quality, and spectrum of clinical utility.Eur. Radiol.20152561598160610.1007/s00330‑014‑3559‑125515204
    [Google Scholar]
  17. ShuHongmin Effects of different ASIR levels and thickness of reconstructive layer on the quality of low-dose CT images of urinary calculi.Acta Universitatis Medicinalis Anhui2019540914891492
    [Google Scholar]
  18. KawashimaH. IchikawaK. HanaokaS. MatsubaraK. TakataT. Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size.J. Appl. Clin. Med. Phys.201819424625110.1002/acm2.1234029729075
    [Google Scholar]
  19. FunamaY. SugayaY. MiyazakiO. UtsunomiyaD. YamashitaY. AwaiK. Automatic exposure control at MDCT based on the contrast-to-noise ratio: Theoretical background and phantom study.Phys. Med.2013291394710.1016/j.ejmp.2011.11.00422182517
    [Google Scholar]
  20. MacDougallR.D. KleinmanP.L. CallahanM.J. Size-based protocol optimization using automatic tube current modulation and automatic kV selection in computed tomography.J. Appl. Clin. Med. Phys.201617132834110.1120/jacmp.v17i1.575626894344
    [Google Scholar]
  21. McColloughC.H. LengS. YuL. CodyD.D. BooneJ.M. McNitt-GrayM.F. CT dose index and patient dose: They are not the same thing.Radiology2011259231131610.1148/radiol.1110180021502387
    [Google Scholar]
  22. BurtonC. SzczykutowiczT. Evaluation of AAPM Reports 204 and 220: Estimation of effective diameter, water-equivalent diameter, and ellipticity ratios for chest, abdomen, pelvis, and head CT scans.J. Appl. Clin. Med. Phys.20171929178549
    [Google Scholar]
  23. LiX. SameiE. SegarsW.P. SturgeonG.M. ColsherJ.G. TonchevaG. YoshizumiT.T. FrushD.P. Patient-specific radiation dose and cancer risk estimation in CT: Part II. Application to patients.Med. Phys.201138140841910.1118/1.351586421361209
    [Google Scholar]
  24. FranckC. VandevoordeC. GoethalsI. SmeetsP. AchtenE. VerstraeteK. ThierensH. BacherK. The role of size-specific dose estimate (SSDE) in patient-specific organ dose and cancer risk estimation in paediatric chest and abdominopelvic CT examinations.Eur. Radiol.20162682646265510.1007/s00330‑015‑4091‑726670320
    [Google Scholar]
  25. LedeniusK. GustavssonM. JohanssonS. StålhammarF. WiklundL-M. Thilander-KLANGA. Effect of tube current on diagnostic image quality in paediatric cerebral multidetector CT images.Br. J. Radiol.20098297631332010.1259/bjr/2440435419188246
    [Google Scholar]
  26. TangK. WangL. LiR. LinJ. ZhengX. CaoG. Effect of low tube voltage on image quality, radiation dose, and low-contrast detectability at abdominal multidetector CT: phantom study.J. Biomed. Biotechnol.201220121610.1155/2012/13016922619490
    [Google Scholar]
  27. HagspielK.D. FlorsL. HousseiniA.M. PhullA. Ali AhmadE. BozlarU. NortonP.T. BonattiH.J.R. Pulmonary blood volume imaging with dual-energy computed tomography: Spectrum of findings.Clin. Radiol.2012671697710.1016/j.crad.2010.12.02122070940
    [Google Scholar]
  28. DalahE.Z. ObaideenA. AnamS. AlzimamiK. JambiL.K. BradleyD.A. Gender based lung cancer risks for symptomatic coronary artery disease patients undergone cardiac CT.PLoS One2022174e026560910.1371/journal.pone.026560935404962
    [Google Scholar]
/content/journals/cmir/10.2174/0115734056267251231121064259
Loading
/content/journals/cmir/10.2174/0115734056267251231121064259
Loading

Data & Media loading...

This is a required field
Please enter a valid email address
Approval was a Success
Invalid data
An Error Occurred
Approval was partially successful, following selected items could not be processed due to error
Please enter a valid_number test